A practical troubleshooting guide for UV-Vis problems: noisy baselines, drift, spikes, poor repeatability — with the cell-related fixes you can try before calling service.
| Symptom | Most likely causes | First fix |
|---|---|---|
| Noisy baseline | Lamp ageing, dirty windows, air bubbles in cell | Clean cells; degas sample; check lamp hours |
| Baseline drift over time | Warm-up not complete, temperature change | Warm up 30 min; control lab temperature |
| Absorbance spikes | Air bubbles, particles in beam path, cell not seated | Tap/fill cell; check alignment; filter sample |
| Poor repeatability | Cell orientation, evaporation, dirty windows | Mark orientation; cover cell; clean windows |
| Negative readings | Blank mismatch, dirty reference cell | See negative absorbance guide |
| Flat/low signal | Wrong Z dimension, cell too empty | Check Z dimension; fill above beam |
If the instrument baseline is clean with empty cells but degrades with a specific cell, the cell is the problem — usually a scratched window, etched surface or a failing glue line. Replacement is cheaper than a service call.