A practical troubleshooting guide for UV-Vis problems: noisy baselines, drift, spikes, poor repeatability — with the cell-related fixes you can try before calling service.
| Symptom | Most likely causes | First fix |
|---|---|---|
| Noisy baseline | Lamp ageing, dirty windows, air bubbles in cell | Clean cells; degas sample; check lamp hours |
| Baseline drift over time | Warm-up not complete, temperature change | Warm up 30 min; control lab temperature |
| Absorbance spikes | Air bubbles, particles in beam path, cell not seated | Tap/fill cell; check alignment; filter sample |
| Poor repeatability | Cell orientation, evaporation, dirty windows | Mark orientation; cover cell; clean windows |
| Negative readings | Blank mismatch, dirty reference cell | See negative absorbance guide |
| Flat/low signal | Wrong Z dimension, cell too empty | Check Z dimension; fill above beam |
If the instrument baseline is clean with empty cells but degrades with a specific cell, the cell is the problem — usually a scratched window, etched surface or a failing glue line. Replacement is cheaper than a service call.
Related: Cuvette Care Guide · Path Length Guide · Size Calculator · Path Length Calculator · Brand Alternatives · Contact Scuvette
Need a hand? Our engineers can help you select the right cuvette for your instrument and application — or supply custom optical cells to your drawing. Contact Scuvette →Usually the cell is dirty, scratched or has fingerprints/oil on the windows. Re-clean the cuvette, handle it by the matte side, and verify the reference (blank) is correct and matched to the sample cell.
Common causes are air bubbles in the beam path, a smudged or scratched window, wrong path length, or using a glass cell in the UV range. Degas the sample, re-clean the cell, and confirm you are using a quartz cell for UV measurements.
Run a quick cell check: swap in a known-good reference cell with the same blank, compare the same sample in two identical cells, and inspect windows under light. If the fault follows the cell, it is a cell issue; if it persists with a clean cell, check the lamp, wavelength calibration and cuvette holder alignment.
Only after thorough cleaning. Precipitates (salts, proteins) leave films that raise baseline absorbance. Soak and rinse until the windows are clear under strong light; if etching or scratches remain, retire the cell from quantitative work.
Negative absorbance means the sample transmits more than the reference. This happens when the blank and sample are swapped, the reference cell is dirtier than the sample cell, or the cells are not matched. Re-run with a correctly prepared blank in the reference position.