Cylindrical Quartz Cuvettes for Turbidity & Scattering

Cylindrical cuvettes are preferred for turbidity and nephelometric measurements because the round walls minimise stray-light artefacts that a flat square face can cause. Fused silica keeps them usable across UV and visible ranges.

Specifications

Outer diameter (OD)22 mm (standard); custom on request
Inner diameter (ID)20 mm (standard)
Path length10 / 20 / 50 mm (standard); custom on request
Volume3.5 / 7 / 17.5 mL (typical for 10/20/50 mm cells)
BottomFlat (standard); round-bottom available on request
MaterialJGS1 fused silica (quartz)

Turbidity & Nephelometry Use

In nephelometry the detector reads light scattered at 90°, and a cylindrical cell gives more uniform scatter geometry than a square cell. This makes cylindrical cuvettes the standard for turbidity and haze measurements.

Choosing a Cylindrical Cell

Match the outer diameter to your instrument's cell holder and the path length to your concentration range. Flat-bottom versions suit stirred or settled samples; round-bottom versions improve mixing.

Frequently Asked Questions

Cylindrical vs square cuvette for turbidity?

Cylindrical cells are preferred for turbidity and nephelometry because the round wall reduces stray-light errors that a flat face can introduce. Square cells are better for standard absorbance in rectangular beams.

Do cylindrical cuvettes work in the UV?

Yes, when made from fused silica (quartz) like these JGS1 cells. Glass cylindrical cells would be limited to the visible range.