A practical cleaning protocol for quartz cuvettes — rinsing, solvent steps, ultrasonic baths, drying and storage — plus what to avoid to keep optical windows scratch-free.
Never wipe a dry cuvette. Dust and particles act as abrasive grit — wiping a dry window scratches it. Always rinse before you touch the optical faces.
Ultrasonic baths are effective for glued (cemented) cells only when kept short and cool: 1–2 minutes in room-temperature mild detergent or water. Prolonged ultrasonic agitation can weaken the glue bond. For one-piece sintered or moulded cells, longer ultrasonic cycles are fine.
| Residue | Solution | Caution |
|---|---|---|
| Protein film | Detergent (0.5%) soak 10 min, then rinse | Avoid strong alkali on glued cells |
| Dyes/pigments | Ethanol or acetone rinse | Check solvent compatibility first |
| Inorganic salts | Dilute acid (e.g. 1% HCl) soak | Rinse thoroughly; no HF ever |
| Oils/grease | Isopropanol or laboratory detergent | Follow with water rinse |
Store clean, dry cells in their plastic box, window surfaces protected, away from dust. For long-term storage, fill nothing — keep them dry and capped.
If the cell still shows haze, scratches or a bad baseline after thorough cleaning, the window surface may be damaged. Matched pairs should be re-verified: see the Cuvette Care Guide for performance verification.