Why your spectrophotometer shows negative absorbance readings — baseline drift, reference mismatch, dirty windows, wrong cell pairing — and how to fix each cause.
Absorbance is defined as log10(I0/I). If the sample transmits more light than the reference — I > I0 — the log goes negative. That usually means the reference/blank is absorbing more than the sample, or the baseline shifted mid-run.
| Cause | Symptom | Fix |
|---|---|---|
| Baseline drift | Slightly negative across whole scan | Re-zero with both cells empty; check lamp warm-up (15–30 min) |
| Blank absorbing more than sample | Negative at specific wavelengths | Blank contaminated or wrong solvent — prepare fresh blank |
| Dirty or mismatched reference cell | Consistent negative offset | Clean both cells; use matched pair for reference |
| Scratched/etched windows | Negative or noisy at UV end | Replace the cell; verify with clean reference |
| Wavelength or bandwidth error | Negative near steep peaks | Validate wavelength calibration |
| Solvent absorbance | Negative where solvent absorbs | Use spectroscopic-grade solvent; correct baseline |
Related: Cuvette Care Guide · Path Length Guide · Size Calculator · Path Length Calculator · Brand Alternatives · Contact Scuvette
Need a hand? Our engineers can help you select the right cuvette for your instrument and application — or supply custom optical cells to your drawing. Contact Scuvette →