Cuvettes en quartz pour le dosage des protéines : A280, Bradford et BCA

Le dosage des protéines s'effectue à deux longueurs d'onde très différentes, et ce seul fait détermine la cuvette nécessaire. L'absorbance UV directe à 280 nm exige le quartz : ni le plastique ni le verre optique ne transmettent à cette longueur. Bradford et BCA lisent dans le visible (562–595 nm), où le verre et le plastique jetable conviennent également. Ce guide couvre les deux cas.

1. Pourquoi l'A280 exige le quartz

Proteins absorb UV at 280 nm through their aromatic residues — tryptophan and tyrosine primarily, with a smaller contribution from phenylalanine and disulfide bonds. Measuring there directly is fast, reagent-free, and the sample is recoverable, which matters when the protein is the product and not just the analyte.

But 280 nm is deep UV. Disposable polystyrene cuvettes cut off around 340 nm and optical glass around 340 nm as well — both are opaque at 280 nm. Only fused silica transmits there. JGS1 synthetic fused silica covers 200–2,500 nm, so a quartz cell handles A280 and every visible assay on the same instrument.

2. UV direct (A280) vs dosages colorimétriques

MethodWavelengthCuvette materialSample recoverable?
A280 direct UV280 nmQuartz (required)Yes
Bradford595 nmQuartz, glass or plasticNo
BCA562 nmQuartz, glass or plasticNo
Lowry650–750 nmQuartz, glass or plasticNo

3. Choix du trajet optique

Beer-Lambert linearity holds roughly between 0.1 and 1.0 absorbance units. Outside that window your numbers drift, and no amount of cuvette quality fixes it.

4. Volume : cuvettes micro et sub-micro

A standard 10 mm macro cell needs roughly 3.5 mL; a semi-micro around 1–1.5 mL. Purified protein rarely comes in those quantities. Micro and sub-micro cells with a masked (black-walled) chamber bring the working volume down to 50–500 µL while keeping the 10 mm path the instrument expects, because the light beam is narrowed rather than the path shortened.

If your sample volume is the binding constraint rather than concentration, this is the first specification to change.

5. Cuvettes recommandées

6. Remarques pratiques

Besoin d'une cuvette en quartz vérifiée à 280 nm ?
Envoyez-nous votre modèle d'instrument, le volume d'échantillon et la plage de concentration — réponse sous 1 jour ouvré. Contact Scuvette →