Cubetas de cuarzo para cuantificación de proteínas: A280, Bradford y BCA

La cuantificación de proteínas se realiza en dos longitudes de onda muy distintas, y ese solo hecho determina la cubeta necesaria. La absorbancia UV directa a 280 nm requiere cuarzo: ni el plástico ni el vidrio óptico transmiten ahí. Bradford y BCA leen en el visible (562–595 nm), donde también sirven vidrio y plástico desechable. Esta guía cubre ambos casos.

1. Por qué A280 requiere cuarzo

Proteins absorb UV at 280 nm through their aromatic residues — tryptophan and tyrosine primarily, with a smaller contribution from phenylalanine and disulfide bonds. Measuring there directly is fast, reagent-free, and the sample is recoverable, which matters when the protein is the product and not just the analyte.

But 280 nm is deep UV. Disposable polystyrene cuvettes cut off around 340 nm and optical glass around 340 nm as well — both are opaque at 280 nm. Only fused silica transmits there. JGS1 synthetic fused silica covers 200–2,500 nm, so a quartz cell handles A280 and every visible assay on the same instrument.

2. UV directo (A280) frente a ensayos colorimétricos

MethodWavelengthCuvette materialSample recoverable?
A280 direct UV280 nmQuartz (required)Yes
Bradford595 nmQuartz, glass or plasticNo
BCA562 nmQuartz, glass or plasticNo
Lowry650–750 nmQuartz, glass or plasticNo

3. Elección del paso óptico

Beer-Lambert linearity holds roughly between 0.1 and 1.0 absorbance units. Outside that window your numbers drift, and no amount of cuvette quality fixes it.

4. Volumen: cubetas micro y sub-micro

A standard 10 mm macro cell needs roughly 3.5 mL; a semi-micro around 1–1.5 mL. Purified protein rarely comes in those quantities. Micro and sub-micro cells with a masked (black-walled) chamber bring the working volume down to 50–500 µL while keeping the 10 mm path the instrument expects, because the light beam is narrowed rather than the path shortened.

If your sample volume is the binding constraint rather than concentration, this is the first specification to change.

5. Cubetas recomendadas

6. Notas prácticas