Quarzküvetten zur Proteinquantifizierung: A280, Bradford und BCA

Die Proteinquantifizierung arbeitet bei zwei sehr unterschiedlichen Wellenlängen, und allein diese Tatsache bestimmt die benötigte Küvette. Direkte UV-Absorption bei 280 nm erfordert Quarz – weder Kunststoff noch optisches Glas transmitieren dort. Bradford und BCA messen im Sichtbaren (562–595 nm), wo auch Glas und Einwegkunststoff funktionieren. Diese Anleitung behandelt beide Fälle.

1. Warum A280 Quarz benötigt

Proteins absorb UV at 280 nm through their aromatic residues — tryptophan and tyrosine primarily, with a smaller contribution from phenylalanine and disulfide bonds. Measuring there directly is fast, reagent-free, and the sample is recoverable, which matters when the protein is the product and not just the analyte.

But 280 nm is deep UV. Disposable polystyrene cuvettes cut off around 340 nm and optical glass around 340 nm as well — both are opaque at 280 nm. Only fused silica transmits there. JGS1 synthetic fused silica covers 200–2,500 nm, so a quartz cell handles A280 and every visible assay on the same instrument.

2. Direktes UV (A280) vs. kolorimetrische Assays

MethodWavelengthCuvette materialSample recoverable?
A280 direct UV280 nmQuartz (required)Yes
Bradford595 nmQuartz, glass or plasticNo
BCA562 nmQuartz, glass or plasticNo
Lowry650–750 nmQuartz, glass or plasticNo

3. Wahl der Schichtdicke

Beer-Lambert linearity holds roughly between 0.1 and 1.0 absorbance units. Outside that window your numbers drift, and no amount of cuvette quality fixes it.

4. Volumen: Mikro- und Submikroküvetten

A standard 10 mm macro cell needs roughly 3.5 mL; a semi-micro around 1–1.5 mL. Purified protein rarely comes in those quantities. Micro and sub-micro cells with a masked (black-walled) chamber bring the working volume down to 50–500 µL while keeping the 10 mm path the instrument expects, because the light beam is narrowed rather than the path shortened.

If your sample volume is the binding constraint rather than concentration, this is the first specification to change.

5. Empfohlene Küvetten

6. Praktische Hinweise

Benötigen Sie eine bei 280 nm geprüfte Quarzküvette?
Senden Sie uns Instrumentenmodell, Probenvolumen und Konzentrationsbereich – wir antworten innerhalb eines Werktages. Contact Scuvette →

Related: Cleaning Protocol · UV-Vis Troubleshooting · Solvent Compatibility · Path Length Guide · Path Length Calculator · Cuvette Care Guide