University procurement faces a split requirement that vendors rarely separate: teaching labs need cells that survive hundreds of inexperienced users at the lowest possible unit cost, while research groups need optical precision, UV transmission and small working volumes. Buying one type for both means overpaying for teaching and under-specifying research. This guide separates the two and gives a procurement checklist for each.Teaching cells get dropped, wiped dry, left full overnight and cleaned with whatever is on the bench. The priorities are therefore:
Research work reaches into the UV, uses small volumes, and often requires fluorescence or flow measurement. The specifications that matter:
| Material | Usable range | Best for | Relative cost |
|---|---|---|---|
| JGS1 fused silica (quartz) | 200 – 2,500 nm | UV, DNA/RNA, protein A280, fluorescence | Highest, but reusable |
| Optical glass | ~340 – 2,500 nm | Visible assays, teaching practicals | Low |
| Disposable plastic | ~340 – 900 nm (typical) | High-throughput visible assays, teaching | Lowest per unit |
The decisive question is simple: does any assay go below 340 nm? If yes, quartz is mandatory and the other two are not options. If no, glass or plastic will save real money.
| Assay / application | Wavelength | Recommended cell |
|---|---|---|
| Nucleic acid quantification | 260 nm | Quartz micro, 10 mm |
| Protein A280 | 280 nm | Quartz macro, 10 mm |
| Bradford assay | 595 nm | Glass or quartz |
| BCA assay | 562 nm | Glass or quartz |
| Fluorescence | Ex/Em dependent | Four-sided polished quartz |
| Kinetics / flow | Varies | Flow cells |
| Concentrated samples | Varies | Short path cells |
For a department equipping one teaching lab and one shared research instrument:
Adjust volumes upward if teaching cohorts are large — breakage, not wear, is what consumes teaching stock.
Related: Cleaning Protocol · UV-Vis Troubleshooting · Solvent Compatibility · Path Length Guide · Path Length Calculator · Cuvette Care Guide